Toshiba TIC-LF232A Manual De Usuario

Descargar
Página de 8
TIC-LF232A 
 
 
2
 
Power supply:  
One of the following can be selected 
100 to 240 Vac  
(Allowable voltage 80 to 264 Vac, 50/60Hz)*
1
 
24 Vdc 
   (Allowable voltage 21 to 27 Vdc)*
2
 
*1
 When the 7
th
 digit of specification code is “B”, 
power supply is 100 to 120Vac  
(Allowable voltage range is 80 to 132VAC, 50/60Hz) 
*2 When the 7
th
 digit of specification code is “B”, 
24Vdc cannot be selected.
 
Power consumption: 
17W (27VA) or less  
(
When 7
th
 digit of specification code is “A”
35W (120VA) or less  
  
(When 7
th
 digit of specification code is “B”)
 
 
Converter Specifications 
Input signals 
Analog signal  the voltage signal from detector, 
proportional to process flow rate. 
Digital input DI (opt.) 
(added when 9
th
 digit of specification code is ”3”) 
Voltage level: High level 20 to 30 V dc 
Low level 2 V dc or less 
Input resistance: 2.7 k
Ω 
Number of inputs: 2 points (DI1 and DI2)
 
DI functions:  
For digital input function, either of the following 
functions can be selected. 
• Range switching  
One Digital Input is used: switching between large 
and small ranges of forward/reverse, 2-range 
measurement. 
Two Digital Inputs are used: switching between 
ranges of single direction, 4-range measurement. 
• Totalizer control input 
  
Totalizer Start/Stop control or Reset/Start 
•Output hold input  
Fixed with set value for current output and pulse 
output (loop check) 
• Zero adjustment  
Starts zero adjustment (on-stream at zero flow rate) 
when DI voltage level goes low after remaining 
high for 10 to 20 seconds. 
Output signals 
Current output: 
4–20mAdc (load resistance 0 to 1K
Ω) 
Digital output (DO) — One point (std.) and three 
more point is optionally available as follows. 
Digital output DO1 (std.):  
Output type: Transistor open collector   
Number of outputs: One point 
Output capacity   : 30Vdc, 200mA maximum 
Digital output DO2DO4 (opt.)  
(
Added when 9
th
 digit of specification code is “3”)  
Out put type: Solid-state relay output (non polarity) 
Number of outputs: 3 points 
Output capacity: 150Vdc, 150mA maximum or 
150Vac (peak to peak),  
100mA maximum 
DO functions—Four digital output function, either 
of the following functions can be selected. 
• Totalizer pulse output  
DO1 or DO2 can be selected (Forward direction 
pulse and reverse direction pulse can be assigned 
independently) 
In the case of DO1  
Pulse rate: 3.6 to 3,600,000 pulses/h 
Pulse width:Settable within the range of  0.3 to 
500ms 
In the case of DO2 
Pulse rate: 3.6 to 360,000 pulses/h 
Pulse width: Settable within the range of 4 to 
500ms 
• Rage switching output 
One Digital Output (DO) is used 
・Single direction, 2-range switching signal 
・Forward/reverse direction switching signal 
Two Digital Outputs (DO) are used 
・Single direction, 4-range switching signals 
・Forward/reverse 2-range switching signals 
• High/low limit alarm output 
An alarm is output when flow rate goes above or 
below the set-point value 
Setting range: -10 to 110% of the settable 
maximum range 
High limit 2 points, low limit 2 points can be set 
At the time of alarm output, Normally Open or 
Normally Closed contact can be selected 
• Preset point output 
Contact ON when totalizer count exceeds the set 
value 
Setting range: 1 to 99999999 count 
• Converter error alarm 
An alarm is output when an operation error is 
detected by self-diagnosis. 
At the time of alarm output, Normally Open or 
Normally Closed contact can be selected