MR850 Technical Manual   *   Revision J   * Issued November 2005   *   Ref.  185041340 (185041713 internal)
 
11
Non-heater wire operation: 
Invasive Mode: Airway temperature < 29.5 °C causes an audible 
and visible alarm. 
Non-invasive Mode: Airway temperature < 26.0 °C causes an 
audible and visible alarm. 
 
Sound Pressure Level:  
Alarms exceed 50 dBA @ 1 m. 
3.4 Performance 
Invasive Mode:  
Flow up to 60 LPM, humidity output >33 mg/L 
Non-Invasive Mode:  
Flow up to 120 LPM, humidity output >10 mg/L 
 
NOTE: Performance results with RT100 breathing circuit 
Maximum System 
Operating Pressure: 
20 kPa, gas leakage at max. pressure <100 mL/minute. 
Warm-up time:  
Less than 30 minutes. 
Recommended ambient 
temperature range: 
18 to 26 °C 
CAUTION:  If operating outside the recommended ambient temperature range, consult your 
local Fisher & Paykel Healthcare representative or refer section 4.1.5. 
3.5 
Transport and Storage Environmental Conditions 
Transport Conditions:  
-10 to +50 °C 
Storage Conditions:    
-10 to +50 °C 
3.6 
Standards and Approvals 
AS/NZS 3200.1.0, CAN/CSA 22.2 No.601.1, UL 60601-1, IEC 60601-1, EN 60601-1