MR850 Technical Manual   *   Revision J   * Issued November 2005   *   Ref.  185041340 (185041713 internal)
 
12 
Operating Modes and Controls 
4.1 Humidifier 
Operation 
The MR850 humidifier is designed to add heat and moisture to respiratory gases.  The gas is 
passed through a humidification chamber where it is warmed and humidified. 
The MR850 has two heating systems.  The first is a heater plate, which heats the water 
contained in the humidification chamber, humidifying the air passing through it.  The humidifier 
monitors the temperature of the gas at the chamber outlet with the chamber probe, and controls 
the amount of power delivered to the heater plate, in order to maintain the chamber set point.  
Under normal conditions the gas is heated to 37 °C in the invasive mode, 31 °C for the non-
invasive mode. 
4.1.1 
Heater wire operation 
 
Figure 4.1 Typical Heater Wire Humidifier Setup 
Humidified gas from the chamber travels through the inspiratory limb, where its temperature 
must be maintained in order to prevent the generated humidity from condensing.  This is 
achieved with a heater wire encapsulated within the inspiratory limb.  The humidifier maintains 
the temperature along the inspiratory limb by monitoring the temperature at the airway probe 
and controlling the power delivered to the heater wire.  Under normal conditions the gas is 
heated to 40 °C in the invasive mode, 34 °C for the non-invasive mode. 
An optional, second heater wire, located in the expiratory limb, minimises condensate in this 
limb. 
Temperature Probe 
RT100 Breathing Circuit 
Ventilator/Gas Supply 
MR850 Humidifier 
MR290 Chamber 
Chamber Probe 
Airway Probe 
Heater Wire Adapter