BAROSTIM NEO LEGACY REFERENCE GUIDE 
9-2
 
Implantation of Replacement IPG 
NOTE:  Verify replacement IPG is compatible with the lead system currently implanted in the patient.  The 
Neo Legacy IPG Model 2100 is compatible with Lead models 101x and can be  used to replace Rheos IPG 
model 2000.  
Surgical approach and techniques for implantation will vary with the preference of the implanting surgeon. 
Essential requirements for proper and safe implantation must include the items covered in this section. 
1.  Connect the CSL to the IPG 
NOTE:  Left lead connects to top header port and Right lead connects to bottom header port of the 
IPG. 
•  Expose the lead terminal. Clean any blood or tissue from the lead terminal and inspect lead 
for any damage prior to attaching it into the IPG. 
•  Insert terminal into appropriate header port.  It may be helpful to place the torque wrench into 
the seal plug prior to inserting the terminal. 
•  Visually verify that the terminal is fully inserted into header by viewing seals visible between 
•  For the port with a lead inserted, use a torque wrench to tighten each set screw in clockwise 
direction until torque wrench begins clicking.   The IPG contains 2 setscrews for each lead 
port; when connecting a therapy lead ensure both are tightened. 
•  Verify proper connection using a slight tug on the lead terminal and an impedance check. 
NOTE:  Electrical connection to the IPG is not established until the setscrew(s) are completely 
tightened using the torque wrench. Do not attempt to deliver any therapy until the connections are 
secured using the torque wrench. 
 
 
                   Right Lead Port 
Figure 8: Terminal insertion into IPG header (Correct insertion) 
 
Both terminal seals located in 
viewing zone (both upper and 
lower header ports)