BAROSTIM NEO LEGACY REFERENCE GUIDE 
9-3
 
 
Figure 9: Terminal insertion into IPG header (Incorrect insertion) 
2.  If only one lead is to be connected, it is recommended that a port plug be placed into the unused lead port 
on the header.  The tip set screw should be tightened when using the port plug.  
3.  If the pocket is in the subcutaneous location place two sutures into the fascia, spaced appropriately for the 
suture holes in the IPG.  This should be a nonabsorbable suture 0 or 1-0 (Ethibond, Silk or Prolene).  
4.  Place the sutures through the suture holes in the IPG header. 
5.  Insert the IPG into the pocket. 
6.  Gently coil excess lead body and place adjacent to the IPG (Figure 10: Correct placement of excess lead 
body) such that the excess lead body is not placed directly in front or behind the IPG (Figure 11: Incorrect 
placement of excess lead body)
.  Ensure that lead body is not pulled taut and allow slack in the path 
between electrode and IPG pocket. 
    
   
Figure 10: Correct placement of excess lead body 
Only one terminal seal in 
viewing zone (both upper and 
lower header ports)