BIOTRONIK SE & Co. KG ICSPGH Manual De Usuario
42
Electromagnetic Compatibility
Electromagnetic Compatibility in
Compliance with
EN 60601-1-2:2002
Compliance with
EN 60601-1-2:2002
•
As the user, you must ensure that the ICS 3000 is
operated in a suitable electromagnetic environment.
operated in a suitable electromagnetic environment.
The following guidelines may not be applicable in all
cases. The propagation of electromagnetic values is,
for example, affected by absorption and reflection by
structures, objects and people. This data is for your
personal information.
cases. The propagation of electromagnetic values is,
for example, affected by absorption and reflection by
structures, objects and people. This data is for your
personal information.
•
The ICS 3000 should not be operated in the vicinity of
devices that display the symbol “Beware of non-
ionizing radiation.” Interference is possible in the
vicinity of such devices.
devices that display the symbol “Beware of non-
ionizing radiation.” Interference is possible in the
vicinity of such devices.
Electromagnetic Emissions (Table 201)
Measuring the
Emitted Interference
Emitted Interference
Compliance
Guidelines for the Electromagnetic
Environment
Environment
High-frequency
emitted interference
emitted interference
According to CISPR 11
Group 1
The device uses RF energy only for its
internal function. Therefore, the emitted
interference of high-frequency waves is
very low and not likely to cause any
interference in nearby electronic
equipment.
internal function. Therefore, the emitted
interference of high-frequency waves is
very low and not likely to cause any
interference in nearby electronic
equipment.
High-frequency
emitted interference
emitted interference
According to CISPR 11
Class B
The device is suitable for use in all
establishments. This includes residences
and facilities directly connected to the
public power supply network that supplies
buildings used for domestic purposes.
establishments. This includes residences
and facilities directly connected to the
public power supply network that supplies
buildings used for domestic purposes.
Emitted interference of
harmonic oscillations
harmonic oscillations
According to
IEC 61000-3-2
IEC 61000-3-2
Class A
Emitted interference of
voltage fluctuations
voltage fluctuations
According to
IEC 61000-3-3
IEC 61000-3-3
Complies