Electromagnetic Compatibility
Electromagnetic Compatibility in 
Compliance with 
EN 60601-1-2:2002
As the user, you must ensure that the ICS 3000 is 
operated in a suitable electromagnetic environment. 
The following guidelines may not be applicable in all 
cases. The propagation of electromagnetic values is, 
for example, affected by absorption and reflection by 
structures, objects and people. This data is for your 
personal information.
The ICS 3000 should not be operated in the vicinity of 
devices that display the symbol “Beware of non-
ionizing radiation.” Interference is possible in the 
vicinity of such devices.
Electromagnetic Emissions (Table 201)
Measuring the 
Emitted Interference
Compliance
Guidelines for the Electromagnetic 
Environment
High-frequency 
emitted interference
According to CISPR 11
Group 1
The device  uses RF energy only for its 
internal function. Therefore, the emitted 
interference of high-frequency waves is 
very low and not likely to cause any 
interference in nearby electronic 
equipment.
High-frequency 
emitted interference
According to CISPR 11
Class B 
The device is suitable for use in all 
establishments. This includes residences 
and facilities directly connected to the 
public power supply network that supplies 
buildings used for domestic purposes.
Emitted interference of 
harmonic oscillations
According to 
IEC 61000-3-2
Class A
Emitted interference of 
voltage fluctuations
According to 
IEC 61000-3-3
Complies