BIOTRONIK SE & Co. KG ICSPGH Manual De Usuario

Descargar
Página de 53
43
Electromagnetic Compatibility
Recommended Safety Distances (Table 206)
Safety distances help prevent interference if you 
maintain a minimum distance between transmitters 
such as mobile RF telecommunication devices and the 
ICS 3000. The necessary distance depends on the 
respective power output of the transmitter.
Note:
At 80 MHz and at 800 MHz, the higher frequency range 
applies.
For transmitters whose maximum output power is not 
indicated in the table, the recommended safety 
distance d can be calculated in meters using an 
equation that is suitable for the respective 
transmission frequency range. P is the maximum 
output power of the transmitter in watts [W] according 
to the specification of the transmitter’s manufacturer.
Transmission 
Frequency
150 kHz to 80 MHz 80 MHz up to 
800 MHz
800 MHz to 
2.5 GHz
Maximum output 
power of the 
transmitter [W]
Safety distance [m]
0,01
0,12
0,12
0,24
0,1
0,37
0,37
0,74
1
1,17
1,17
2,34
10
3,70
3,70
7,40
100
11,7
11,7
23,4
Transmission 
Frequency
150 kHz to 80 MHz 80 MHz up to 
800 MHz
800 MHz to 
2.5 GHz
Equation
d
1 17 P
,
=
d
1 17 P
,
=
d
2 34 P
,
=