Cleveland Medical Devices Inc. 
                                                             BioRadio 110 User Guide
 
                                                                                                                                                  43 
Chapter 6:  Troubleshooting & Frequently Asked 
 
      
    
            Questions 
If your BioRadio is not performing properly, use this troubleshooting chart for quick so-
lutions to common problems.  If the problem persists, call Cleveland Medical Devices 
Problem... 
Possible Cause... 
Remedy... 
No data is displayed  Receiver Power cord or battery 
Connect power cord or battery 
 
Transmitter power switch off? 
Turn on power. 
 
Faulty power outlet? 
Try another power outlet. 
 
Receiver cable is not connected 
properly to PC? 
Connect Receiver cable to 
RS232 port properly. Check 
 
Transmitter is not connected 
Connect transmitter properly. 
 
Transmitter batteries are dead.  
Receiver battery is dead. 
Replace batteries making sure 
the poles are in the proper 
Signal image breaks 
up 
Dropped packets? 
Bring the Receiver and 
Transmitter closer together.  
Try not to block the 
transmission path with the 
subject.  Move large metal 
Signal image is cut- Amplitude of Data outside visible  Adjust Scale Control. 
Signals are hard to 
visualize 
The channel color selected does 
not contrast well with the 
Change channel color. 
Computer freezes up 
or display graph 
PC needs a faster processor 
Run BioRadio Capture on a 
Pentium™ 200/MMX or 