After starting Capture Mode, no data is seen in the Display Graph, and the
status bar shows 0% packets received.
status bar shows 0% packets received.
Possible cause: The Transmitter is not turned on.
Solution:
Solution:
Turn the power switch on the Transmitter to the "On"
position.
Possible cause: The Transmitter batteries are dead.
Solution:
Replace the Transmitter batteries.
Possible cause:
The Transmitter and Receiver are set to different RF
frequencies.
Solution:
Call the Toll-free Customer Service Number to request an
RMA number and return the device so that the frequency
can be reprogrammed.
can be reprogrammed.
The channel signal does not appear in the Display Graph.
Possible cause: The signal is outside the viewing scale.
Solution:
Solution:
Adjust the channel's Scale Control until the signal is visible.
Possible cause: The selected channel color does not contrast well with the
graph background.
Solution:
Change the channel's display color. (See Channels
Dialog).
The Display Graph appears "choppy" or sluggish while displaying data in
Capture Mode.
Capture Mode.
Possible cause: The computer is too slow to display the graph data.
Solution:
Solution:
Run BioRadio Capture on a Pentium™ 200/MMX or faster
computer.
Possible cause:
The Timescale is too large to display the requested number
of data points.
Solution:
Change the timescale to a smaller value. (See Timescale
Dialog).
Possible cause:
The Display Resolution is configured too high for the
computer.