BioRadio 110 User Guide 
                                                             Cleveland Medical Devices Inc.
44 
When starting Capture Mode, an error message is displayed stating that 
the selected COM port is in use. 
 
Possible cause:
  Another application is using the specified COM port. 
Solution: 
             Close the other application, or use a different COM port.  
(See Capture Options). 
 
Possible cause:   
     Your computer does not support the specified COM port. 
Solution:   
           Use a different COM port.  (See Capture Options). 
 
When starting Capture Mode, an error message is displayed stating that 
no BioRadio Receiver could be found on the specified COM port. 
 
Possible cause:
   
     The Receiver cable is attached to the wrong COM port. 
Solution:  
            Attach the Receiver cable to the correct COM port. 
 
Possible cause:
  BioRadio Capture is configured to use the wrong COM port. 
Solution: 
             Configure BioRadio Capture to use the correct COM port.  
(See Capture Options). 
 
Possible cause:  Power is not applied to the Receiver. 
Solution: 
             Check that the power cable or battery is properly connected 
to the Receiver and that your power source works. 
 
After starting Capture Mode, the status bar shows less than 100% packets 
received. 
 
Possible cause:
  The Transmitter is out of range. 
Solution: 
             Keep the Transmitter within range of the Receiver (within 25 
feet indoors, or up to 100 feet line-of-sight). 
 
Possible cause:  The Transmitter batteries are weak. 
Solution: 
             Replace the Transmitter batteries (new AAA batteries should 
last approximately 10-12 hours). 
392-0000rev1.pub 
page 44