TransCore IT241005422 Manual De Usuario
IT2410 Tag Programmer User Guide
4-6
Programming Cycle
When the programming cycle is initiated by the remote host system, the PROGRAM
LED illuminates and the alarm beeps for an approximate duration of 0.25 seconds,
once a second, for a 10-s period, or until the programming operation has completed.
LED illuminates and the alarm beeps for an approximate duration of 0.25 seconds,
once a second, for a 10-s period, or until the programming operation has completed.
The operator has an approximate 10-s time-out period in which to place an operating
IT2200-series tag on the programming head fixture. Once the tag is in the fixture and
the programmer detects the valid RF data, the programmer executes the specific write
command from the host.
IT2200-series tag on the programming head fixture. Once the tag is in the fixture and
the programmer detects the valid RF data, the programmer executes the specific write
command from the host.
If the programming is successful then the PROGRAM LED goes out and the alarm
sounds three times for approximately 0.25 seconds each.
sounds three times for approximately 0.25 seconds each.
If the programming cycle is unsuccessful or a tag is not detected within the 10-s
period, the ERROR LED illuminates and the alarm sounds once for approximately 1
second.
period, the ERROR LED illuminates and the alarm sounds once for approximately 1
second.
Verifying Cycle
When the verify cycle is initiated by the remote host system, the VERIFY LED illumi-
nates and the alarm beeps each second for an approximate 10-s period or until the ver-
ify operation is complete.
nates and the alarm beeps each second for an approximate 10-s period or until the ver-
ify operation is complete.
The operator has a 10-s time-out period in which to place an operating IT2200-series
tag on the programming head. Once the tag is placed in the fixture and the program-
mer can detect the valid RF data, the programmer executes the specific read command
from the host.
tag on the programming head. Once the tag is placed in the fixture and the program-
mer can detect the valid RF data, the programmer executes the specific read command
from the host.
If the verifying is successful the VERIFY LED goes out and the alarm sounds three
short beeps.
short beeps.
If the verifying cycle is unsuccessful or if a tag is not detected within the 10-s period,
the ERROR LED illuminates and the alarm beeps for 1 second.
the ERROR LED illuminates and the alarm beeps for 1 second.
The tag programmer incorporates an audible indicator that is controlled by software.
The indicator’s sound is pleasing to the ear (2200 Hz) and can be easily heard while
operating near the unit under normal conditions. The alarm audibly indicates the status
of programmer operations. Table 4-1 lists the various programmer operations and the
corresponding alarms that sound as an audible operation verification.
The indicator’s sound is pleasing to the ear (2200 Hz) and can be easily heard while
operating near the unit under normal conditions. The alarm audibly indicates the status
of programmer operations. Table 4-1 lists the various programmer operations and the
corresponding alarms that sound as an audible operation verification.
Table 4-1 Programmer Operations and Corresponding Alarms
Programmer Operation
Audible Alarm
Beep Duration
Duty (%)
Successful Self Test
3 beeps
0.25 s
50
Failed Self Test
1 beep
3.0 s
100