Discharge Tab
Patient Setup Functions
5 - 38
0070-10-0690-01
Panorama™ Operating Instructions
5.3.4
Discharging a Patient from a Telepack Device 
The same messages described in “Discharging a Patient from the Central Station and the 
Bedside Monitor” on page 5-36 will 
be displayed, with one exception. If the Yes button is 
selected after the message Are you sure you want to discharge this patient? 
Selecting Yes will move the patient to the discharge list is displayed, the Telepack 
will be discharged from the central station. Discharging a Telepack device can only be done 
from the central station. Once discharged, all the patient information will be cleared.     
5.3.5
Sidebar Button - Normal Screen
Select the Normal Screen sidebar button to return the display to the full screen mode, and 
close all tabs, message boxes, and dialog boxes.
5.3.6
Troubleshooting
This section lists some of the potential messages and issues that may occur while using the 
Discharge tab. 
MESSAGE/ISSUE * 
REASON
SOLUTION
No patient Selected!
This message appears in the 
demographic field when either 
a patient tile has not been 
selected or an empty tile was 
selected.
Select an active patient tile.
The patient’s data and 
submitted print jobs will 
be deleted because at 
least two of the following 
fields: First Name, Last 
Name or ID has not been 
filled in. Do you still want 
to proceed with the 
discharge?
 
The selected patient does not 
have at least two of the three 
demographic fields completed. 
The demographic fields include 
First Name, Last Name, and 
ID.
Select the No button to close 
the dialog, then enter the 
required demographic 
information. 
The patient’s data and 
submitted print jobs will 
be deleted because at 
least two of the following 
fields: First Name, Last 
Name or ID has not been 
filled in. Do you still want 
to proceed with the 
discharge?
 dialog box is 
open and was automatically 
closed without any user 
interaction.
The bedside monitor has gone 
into a Communications Lost 
state. 
Restore the connection 
between the Panorama Central 
Station and the monitoring 
device. If unable to restore the 
connection, select the 
Discharge Patient button to 
discharge the equipment from 
the patient tile.
Unable to perform 
operation on tile: Central 
to Central Transfer Out in 
progress.
A Central to Central transfer 
Out has been initiated at the 
source central station but not 
accepted or rejected at the 
destination central station for 
the selected tile, and the 
Discharge Patient button is 
pressed in the Discharge tab 
at the source central station. 
Wait until the Central to 
Central transfer is complete, 
then discharge the patient. 
*
Messages are shown in all bold text.