Display Tab
Patient Setup Functions
5 - 40
0070-10-0690-01
Panorama™ Operating Instructions
5.4
Display Tab
The Display tab (shown in FIGURE 5-21) is used to select the digital and waveform 
parameters that will appear in the selected patient tile. The default settings for patient tiles 
are defined by the Unit Choices tab, which is described in the “Unit Choices Tab” on 
page 9-50.
 The layout and functionality of the Display tab will change based on the type of 
monitoring device assigned to the selected patient tile. 
The Display tab is available at the ViewStation and the WorkStation (shown in 
FIGURE 5-22), and functions in the same manner as it does at the host central station. When 
a patient tile is assigned at the ViewStation or the WorkStation, it will initially use the display 
settings of the associated patient tile at the host central station, provided the patient tile is in 
an active state.  
After this initial transfer, changes made to the host central station, the ViewStation and/or the 
WorkStation patient tile Display tab will be independent of each other unless 
communication is interrupted. Automatic updating of the display settings at the remote 
workstation patient tile will occur if any of the following occur:
• The monitoring device is placed in the Standby mode at the central station
• The monitoring device is not communicating with its host central station (Communications 
Lost)
• Communication between the host central station and the ViewStation and/or the 
WorkStation is lost (Central Communications Lost)
NOTE:
The host central station(s) C-LAN configuration settings and 
the software installed at the remote location dictates the 
functionality available for each patient tile. For additional 
information, refer to ‘‘The ViewStation and the 
WorkStation’’ on page 2-4.
5.4.1
Accessing the Display Tab
1. From the menu bar, select the Patient Setup menu button. The tabs associated with 
patient setup are displayed.
2. Choose a patient tile by selecting its digital tile or waveform area.
3. Select the Display tab. The Display tab is enabled.
 
FIGURE 5-21 Display Tab