Panorama™ Operating Instructions
0070-10-0690-01
5 - 39
Patient Setup Functions
Discharge Tab
Unable to perform
operation on tile: Central
to Central Transfer In in
progress.
operation on tile: Central
to Central Transfer In in
progress.
A Central to Central transfer In
is in progress for the selected
tile and the selected tile is
actively monitoring a patient,
and the Discharge Patient
button is pressed in the
Discharge tab at the
destination central station.
is in progress for the selected
tile and the selected tile is
actively monitoring a patient,
and the Discharge Patient
button is pressed in the
Discharge tab at the
destination central station.
Wait until the Central to
Central transfer is complete,
then discharge the patient.
Central transfer is complete,
then discharge the patient.
The Discharge Patient
button in the Discharge tab is
disabled.
button in the Discharge tab is
disabled.
The Discharge at Bedside
setting in the Unit Choices
tab is set to OFF.
setting in the Unit Choices
tab is set to OFF.
If a patient discharge is
necessary, it must be
performed from the bedside
monitor.
necessary, it must be
performed from the bedside
monitor.
The Discharge tab needs to
be refreshed.
be refreshed.
Select another tab and then
return to the Discharge tab.
return to the Discharge tab.
No patient is selected.
Select a patient tile.
The Discharge Patient
button in the Discharge tab is
enabled.
button in the Discharge tab is
enabled.
The bedside monitor has gone
into a Communications Lost
state.
into a Communications Lost
state.
Verify the connection between
the Panorama Central Station
and the monitoring device.
the Panorama Central Station
and the monitoring device.
The Discharge at Bedside
setting in the Unit Choices
tab is set to OFF.
setting in the Unit Choices
tab is set to OFF.
No solution required.
A Telepack device is in use.
No solution required.
Local Discharge Only!
Patient will not be
discharged at bedside.
Do you still want to
proceed with discharge?
Patient will not be
discharged at bedside.
Do you still want to
proceed with discharge?
The bedside monitor is in a
Communications Lost state.
Communications Lost state.
Verify the connection between
the Panorama Central Station
and the monitoring device.
the Panorama Central Station
and the monitoring device.
The Local Discharge Only!
Patient will not be
discharged at bedside.
Do you still want to
proceed with discharge?
dialog box is open and was
automatically closed without
any user interaction.
Patient will not be
discharged at bedside.
Do you still want to
proceed with discharge?
dialog box is open and was
automatically closed without
any user interaction.
The beside monitor is no
longer in a Communications
Lost state.
longer in a Communications
Lost state.
If the Discharge at Bedside
button in the Unit Choices
tab is set to ON, proceed with
the discharge by selecting the
Discharge Patient button. If
it is set to OFF, the discharge
must be performed from the
bedside.
button in the Unit Choices
tab is set to ON, proceed with
the discharge by selecting the
Discharge Patient button. If
it is set to OFF, the discharge
must be performed from the
bedside.
MESSAGE/ISSUE *
REASON
SOLUTION
*
Messages are shown in all bold text.