BMC Medical Co. Ltd. YH600AYH600B Manual De Usuario
YH‐600B Pro User Manual V1.0
25 / 49
②
Detach the wrist belt from the main device.
③
İt is important to always hand-wash the wrist belt.
④
Wash the wrist belt in warm (approx. 40°C) and soapy water. Use pure soap.
⑤
Rinse the wrist belt well with potable water and allow it to air dry out of direct sunlight.
CAUTION!
•
Never use abrasive agents, alcohol, chlorine-containing substances, acetone, or other solvents to
clean the device.
•
Do not immerse the device, the airflow kit, or the pulse oximeter sensor in fluids, and ensure that
no fluids penetrate into the product. Do not immerse the main device in water.
•
The Airflow Cannula is non-reusable, dispose of it in accordance with local regulations.
12.2 Disinfection
If necessary, i.e. after infectious diseases, uses between patients or if parts have become unusually
dirty, you can disinfect certain parts.
dirty, you can disinfect certain parts.
Clean YH-600B Pro parts as instructed above. Disinfect YH-600B Pro as follows:
We recommend CaviCide
®
liquid for disinfection.
We recommend you to wear suitable gloves (e.g. house-hold or disposable gloves) during disinfection
procedures.
procedures.
Disinfect the Pulse Oximeter Sensor
The pulse oximeter sensor may be surface-cleaned by wiping it with CaviCide
®
solution.
①
Detach the pulse oximeter sensor from YH-600B Pro main device.
②
Saturate a clean, dry gauze pad with the CaviCide
®
solution. Wipe all surfaces of the pulse oximeter
sensor with this gauze pad.
③
Saturate another clean, dry gauze pad with potable water. Wipe all surfaces of the pulse oximeter
sensor with this gauze pad.
④
Discard the used potable water. Always use fresh volumes of water for each wipe. Do not reuse the
water for wipe or any other purpose. Repeat rinsing the pulse oximeter sensor for two additional times,
for a total of three wipes to remove CaviCide
for a total of three wipes to remove CaviCide
®
solution residues.
⑤
Dry the pulse oximeter sensor by wiping all surfaces with a clean, dry gauze pad.
Disinfect the CPAP Pressure Tube
①
Detach the CPAP pressure tube from the airflow kit.
②
Immerse the CPAP pressure tube completely, filling all lumens and eliminating air pockets, in
CaviCide
®
solution for a minimum of 5 minutes at 20℃(68℉) or higher to destroy all pathogenic
microorganisms.
③
Rinse the CPAP pressure tube with potable water. Keep the CPAP pressure tube totally immersed