for a minimum of 1 minute. Manually flush all lumens with large volumes (not less than 100 ml) of rinse 
water. 
  Remove the CPAP pressure tube and discard the rinse water. Always use fresh volumes of water for 
each rinse. Do not reuse the water for rinsing or any other purpose. Repeat rinsing the CPAP pressure 
tube for two additional times, for a total of three wipes to remove CaviCide
®
 solution residues. 
  Dry the CPAP pressure tube by wiping all surfaces with a clean, dry gauze pad and air dry the lumen 
of the CPAP pressure tube. 
Disinfect the Wrist Belt 
  Detach the wrist belt from the main device. 
  Immerse the wrist belt completely in CaviCide
®
 solution for a minimum of 5 minutes at 20℃(68℉) or 
higher to destroy all pathogenic microorganisms. 
 Rinse the wrist belt with potable water. Keep the wrist belt totally immersed for a minimum of 1 
minute in duration. 
  Remove the wrist belt and discard the rinse water. Always use fresh volumes of water for each rinse. 
Do not reuse the water for rinsing or any other purpose. Repeat rinsing the wrist belt for two additional 
times, for a total of three wipes to remove CaviCide
®
 solution residues. 
  Air dry the wrist belt out of direct sunlight. 
Sterilization of the device and its components other than recommended is not permitted. 
WARNING!       
The device shall not be serviced or maintained while in use with a patient. 
13. Disposal 
Dispose of the device and its package in accordance with local regulations. 
Do not dispose device as household waste. To ensure proper disposal of the device, consult an 
authorized and certified electronic scrap recovery firm. You can find out their address from your 
environmental officer or your municipal authorities.