MEASure<x>
Query
Description
Returns the test parameters & results of the tester
in either MANU or AUTO mode.
MANU mode: Returns the test parameters &
results of a MANU test.
AUTO mode: Returns the test parameters & results
of the selected step (1-16) of the AUTO test.
Return parameters: function, judgment/status, test
voltage, test current/resistance, test time (time of
completed test) or ramp time (elapsed time of test
that has not been completed.
in either MANU or AUTO mode.
MANU mode: Returns the test parameters &
results of a MANU test.
AUTO mode: Returns the test parameters & results
of the selected step (1-16) of the AUTO test.
Return parameters: function, judgment/status, test
voltage, test current/resistance, test time (time of
completed test) or ramp time (elapsed time of test
that has not been completed.
Query Syntax
MEASure<x>?
Parameter
(MANU mode)
(MANU mode)
No parameter needed for MANU
mode.
mode.
Parameter
(AUTO mode)
(AUTO mode)
<x>
<NR1>1~16. Step number.
Return parameter <string>
Returns the test status of the test
in the following format:
function, judgment or status, test
voltage, test current or resistance,
test time or ramp time
in the following format:
function, judgment or status, test
voltage, test current or resistance,
test time or ramp time
Function
ACW, DCW, IR, GB
Judgment
/Status
/Status
PASS, FAIL
VIEW
VIEW
Test voltage
voltage+unit
Test current
/Test resistance
/Test resistance
current+unit
resistance+unit
resistance+unit
Test time
/Ramp time
/Ramp time
T=time+S
R=time+S
R=time+S
Example
(in MANU mode)
(in MANU mode)
MEAS?
>ACW, FAIL , 0.024kV ,0.013 mA ,R=000.1S
Returns the test result of the current manual test.
>ACW, FAIL , 0.024kV ,0.013 mA ,R=000.1S
Returns the test result of the current manual test.