12
5 Maintenance
WARNING
To avoid electrical shock, disconnect the meter from any circuit, remove the test leads
from the input jacks and turn OFF the meter before opening the case. Do not operate
with open case.
Cleaning and Storage
Periodically wipe the case with a damp cloth and mild detergent; do not use abrasives
or solvents. If the meter is not to be used for periods of longer than 60 days, remove the
batteries and store them separately
Trouble Shooting
If the instrument fails to operate, check batteries and test leads etc., and replace as
necessary. Double check operating procedure as described in this user’s manual.
If the instrument voltage-resistance input terminal has subjected to high voltage
transient (caused by lightning or switching surge to the system) by accident or abnormal
conditions of operation, the series fusible resistors will be blown off (become high
impedance) like fuses to protect the user and the instrument. Most measuring functions
through this terminal will then be open circuit. The series fusible resistors and the spark
gaps should then be replaced by qualified technician. Refer to the LIMITED
WARRANTY section for obtaining warranty or repairing service.
Battery and Fuse replacement
Battery use:
Standard 1.5V AAA Size (NEDA 24A or IEC LR03) battery X 2
Standard CE Version MD 9030 series fuses:
Fuse (FS1) for
AmA current input: 0.5A/250V, F fuse;
Fuse (FS2) for A current input: 15A/250V, F fuse
Protection-reinforced CE + UL Version MD 9030 series fuses:
Fuse (FS1) for
Protection-reinforced CE + UL Version MD 9030 series fuses:
Fuse (FS1) for
AmA current input: 0.63A/500V, IR 200kA, F fuse;
Fuse (FS2) for A current input: 12.5A/500V, IR 20kA, F fuse
Battery replacement for models with battery
access door:
Loosen the 2 screws from the battery access
door of the case bottom. Lift the battery access
door and thus the battery compartment up.
Replace the battery. Re-fasten the screws.
Fuse replacement (and also Battery
replacement for splash proof version without
battery access door):
Loosen the 4 screws from the case bottom. Lift
the end of the case bottom nearest the input
jacks until it unsnaps from the case top.
Replace the blown fuse(s) and/or the battery.
Replace the case bottom, and ensure that all
the gaskets are properly seated and the two
snaps on the case top (near the LCD side) are
engaged. Re-fasten the screws..