Metrel MD 9030 Digital-Multimeter, DMM, 20991441 Hoja De Datos
Los códigos de productos
20991441
MD 9030 TRMS Electrical & Electronic DMM with Temperature
Specification
14
Low Battery:
Below approx. 2.4V
E.M.C.:
Meets EN61326(1997, 1998/A1), EN61000-4-2(1995), and EN61000-4-3(1996)
In an RF field of 3V/m:
In an RF field of 3V/m:
Capacitance function is not specified
AC 4.000V range: Total Accuracy = Specified Accuracy + 700 digits
AC
AC
400.0
A range: Total Accuracy = Specified Accuracy + 300 digits
Other function ranges: Total Accuracy = Specified Accuracy + 40 digits
Performance above 3V/m is not specified
Sleep Mode Timing:
Idle for 30 minutes
Sleep Mode Consumption: 360
A typical for MD 9030
Dimension:
L186mm X W87mm X H35.5mm; L198mm X W97mm X H55mm with holster
Weight:
296 gm; 396 gm with holster
Special Features:
25ms Max Hold; Data Hold; Relative zero mode; Beep-jack™ input warning;
Back-lighted display
Back-lighted display
Accessories:
Test leads (pair), batteries installed, user's manual
Optional Accessories: Banana plug type-K bead probe AMD9023 x 1 (Banana pins to type-K socket
plug adapter AMD9024
Electrical Specifications
Accuracy is ±(% reading digits + number of digits) or otherwise specified, at 23
O
C ±5
O
C & less than 75%
R.H. Model MD 9030 True RMS accuracy of ACV & ACA is specified from 5 % (10% for AC400.0mV
range) to 100 % of range, or otherwise specified. Maximum Crest Factor < 1.75 : 1 at full scale & < 3.5 : 1
at half scale, and with frequency components within the specified frequency bandwidth for non-sinusoidal
waveforms
range) to 100 % of range, or otherwise specified. Maximum Crest Factor < 1.75 : 1 at full scale & < 3.5 : 1
at half scale, and with frequency components within the specified frequency bandwidth for non-sinusoidal
waveforms
DC Voltage
RANGE Accuracy
400.0 MV
0.3% + 4D
4.000V, 40.00V,
400.0V
400.0V
0.5% + 3D
1000V
1.0% + 4D
NMRR:>50dB @ 50/60Hz
CMRR:>120dB @ DC, 50/60Hz, Rs=1k
CMRR:>120dB @ DC, 50/60Hz, Rs=1k
Input impedance : 10M
, 30pF nominal
(1000M
for 400.0mV range)
Max Hold (Voltage & Current)
Specified accuracy ± 50 digits for changes > 25ms in duration
AC Voltage
RANGE Accuracy
50HZ -- 500HZ
400.0mV*
400.0mV*
4.0% + 5d
4.000V, 40.00V,
400.0V
400.0V
1.5% + 5d
1000V
4.0% + 5d