Measurement Options 
46 
 
 
Warning:    Readings inside the buffer are volatile.  Therefore do not power off the instrument or 
data will be lost.    Additionally, each time the store reading function is enabled  and initialized, 
previously stored data in the buffer will be overwritten. 
 
4.3.2    Recall Readings 
Use the following steps to recall the stored readings and their statistical information (minimum, maximum, 
average, and standard deviation): 
1.  Press 
  →        to enter the menu on the menu level, “A: MEAS MENU” will be displayed. 
2.  Use        or        key to move across to the SYS MENU on the menu level, “D: SYS MENU” will 
be displayed. 
3.  Press        to move down to the command level within the SYS MENU, “1: RDGS STORE” will be 
displayed. 
4.  Use        or        key to move across to the SAVED RDGS command, “3: SAVED RDGS” will be 
displayed. 
5.  Press        to move down a level to view the stored readings.  “MEM” annunciator  will turn on to 
indicate the instrument accessing buffer memory. If no readings are stored,  a “BUFFER EMPTY” 
message will be displayed for a few seconds. 
6.  Use         or      key to navigate through the readings.  The display may look like any of the 
following when browsing through all the stored readings:   
 
7.  Use the function keys or press 
  →        to exit from the menu and return to measurement 
mode. 
 
Warning:    Readings inside the buffer are volatile.  Therefore do not power off the instrument or 
data will be lost.    Additionally, each time the store reading function is enabled  and initialized, 
previously stored data in the buffer will be overwritten. 
 
1.234    m:005 
Reading 
Unit 
Prefix 
Reading # or type of 
statistical information 
Reading:  Stored reading value 
Unit Prefix:  The unit prefix that applies 
to the Reading.    This can be m (10
-3
), u 
(10
-6
), n (10
-9
), or blank. 
Reading # or type of statistical 
information:  Indicates the reading # or 
count that corresponds to the Reading. 
This can be any number from 001 – 512.  
Additionally, this can display the type of 
statistical information in which the 
Reading corresponds with.    This can be 
MAX, MIN, AVG, or STD (See section 
 for details).