Bacharach PCA 3 Manual De Propietario
PCA
3 Manual
0024-9472 Rev 2
23
CHAPTER 4. CONFIGURATION PROCEDURES
4.1. Default Operating Parameters
The PCA
®
3 is set up at the factory for the following operating parameters. To
change any of these parameters, perform the associated procedure that follows.
Parameter
Factory Default
To Change Value, See…
Fuel
Natural Gas
Auto/Manual CO Zero
Auto Zero
Temperature Units
°F
Pressure Units
Inches of Water Column (inwc)
Pollution Units
ppm
Date
Current MM/DD/YY
Time
Current EST HH:MM AM/PM
O
2
Reference
0%
Print Pressure
No
Zoom
Standard
Logging
No
Button Sound
On
Test ID Information
<blank>
User Name
<blank>
Language
English
Calibration Reminder
Never
Run/Hold Screen
Format
Format
See page 37 for default
4.2. Fuel Selection
To assure the accurate calculation of combustion efficiency, select the fuel being
burned as follows.
burned as follows.
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the
▲▼ buttons to highlight FUEL and then press ENT to display
the FUEL MENU.
3. Use the
▲▼ buttons to scroll through the list of available fuels until the
desired fuel is highlighted. In the example shown, PROPANE has been
selected (If custom fuels are added, they will be displayed at the bottom
of the list).
selected (If custom fuels are added, they will be displayed at the bottom
of the list).
NOTE: Use the
◄► buttons to quickly scroll to the bottom
and top of the list.