NOTE:  The PCA
®
3 performs a zero function during warm-up.  
Among other purposes, an auto-zero determines the “zero reading” 
of the CO channel in fresh air and provides an offset for the CO 
measurement.  A manual zero  detects background CO during 
startup. 
 
 
Manual Zero:
 
  When the CO channel is set to manual zero, the analyzer does not 
zero the CO sensor to ambient conditions during start up.  In this 
mode, the “fresh air zero” established during manual mode setup 
is stored in memory and used for the calculation of CO. 
Auto Zero:
Auto Zero determines the zero reading of the CO channel in fresh air and 
provides an offset for the CO measurements.  Manual zero is used to detect CO 
that may be present during start-up.  The auto/manual zeroing option allows the 
operator to select the start-up mode.  It is selectable from the CO Zero Setting of 
the Setup Menu. 
 
When the CO channel is set to auto zero, the CO sensor is zeroed 
to the ambient CO level during start up. Important:  When using 
this mode, the analyzer must be turned ON in fresh air; otherwise, 
incorrect CO readings will occur. 
 
 
NOTE:  Both methods take the user through a 60-second count 
down to establish a new zero.  The manual mode establishes a 
fresh air zero and stores it for use during instrument startup. 
 
 
 
4.4.  Temperature Units Selection 
 
Select to display temperature in either °F or °C as follows. 
1.  Display the MAIN MENU by pressing the MENU (F2) button. If 
necessary, press ESC until MENU appears above F2
2.  Use the 
▲▼ buttons to highlight SETUP, and then press ENT to display 
the SETUP MENU. 
3.  Use the 
▲▼ buttons to highlight TEMP UNITS, and then press ENT to 
display the TEMP UNITS MENU. 
4.  Use the 
▲▼ buttons to highlight the desired temperature units. In the 
example shown, Fahrenheit has been selected. 
5.  Press ENT to save the selection and re-display the SETUP MENU.