Bacharach PCA 3 Manual De Propietario
PCA
3 Manual
24
0024-9472 Rev 2
4. Press ENT to save the selection and display the Combustion Test
HOLD screen. Observe that the name of the selected fuel should now
appear at the top of the screen.
appear at the top of the screen.
NOTE: In addition to the standard fuels programmed into the
PCA
PCA
®
3, Bacharach can develop custom fuel codes based on the
customers specific needs. The PCA
®
3 can be programmed with 2
additional fuels which can be added to the instrument using the
Fyrite User Software (FUS). Consult factory for price and delivery.
Fyrite User Software (FUS). Consult factory for price and delivery.
Figure 4-1. Fuel Selection
4.3. Auto/Manual CO Zero Selections
Select how the PCA
®
3 performs a zero function.
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the
▲▼ buttons to highlight SETUP, and then press ENT to display
the SETUP MENU.
3. Use the
▲▼ buttons to highlight CO Zero Setting, and then press ENT
to display the CO ZERO SETTING MENU.
4. Use the
▲▼ buttons to highlight the desired zeroing option (Manual or
Auto-Zero). In the example shown, Manual Zero has been selected.
5. Place the instrument in fresh air to zero, and press ENT to start a 60-
second count down.
6. Press ENT to save and re-display the SETUP MENU.
Figure 4-2. Auto/Manual Zero Selection