F-2200 Series Vortex Flow Meter Installation and Operation Guide • Revised 12/08
SECTION 1.0: INTRODUCTION
We, at ONICON INCORPORATED, would like to thank you for purchasing our F-2200 Series Vortex Flow 
Meter. As our valued customer, our commitment to you is to provide fast reliable service and assistance, 
while continuing to offer you new products to meet your growing fl ow measurement needs.
1.1  PURPOSE OF THIS GUIDE
 
The purpose of this guide is to provide installation and commissioning procedures and basic 
 
operating and servicing instructions for ONICON F-2200 Series Vortex Flow Meters.
1.2  PRINCIPAL OF OPERATION
 
ONICON F-2000 series vortex fl ow meters utilize Karman’s vortex street principal  to detect changes 
 
in velocity in the medium fl owing around the titanium shedder bar and through the meter body. A 
 volumetric 
fl ow rate is then derived by utilizing the known cross-sectional area of the meter body 
 
and the average velocity of the medium. When provided with optional temperature and/or pressure 
 
compensation the F-2000 series fl ow meters are capable of calculating the required density 
 
corrections necessary to provide direct mass fl ow measurement of liquids, steam or gases.
1.3  FEATURES AND SPECIFICATIONS
CONSTRUCTION 
 
Meter body: 
 
 
Casting: 316 stainless steel for diameters up to 4”    
 
 
 
304 stainless steel for 6” and 8” diameters 
 
 
Shedder bar:  Unalloyed titanium 
 
 
Shedder bar seal:  Ni plated Inconnel 
 
 
Damper pin o-ring:     
 
 
 
Viton for, non-steam, -4° to 356° F     
 
 
 
Kalrez 4079 for, non-steam, -4° to 428° F     
    
Parofl uor for steam, -4° to 464° F   
 
Electronics enclosure: 
 
 
Enclosure material: Die-cast aluminum 
 
 
Enclosure viewing window: Tempered glass, 
 
 
 
UV blocking  
CALIBRATION 
 
Each meter undergoes a 5 point velocity calibration 
 
from 0-250 ft/sec  
PROGRAMMING 
 
Each meter is programmed at the factory based on application specifi c data provided by the 
 
customer. Field re-programmable via keypad, with coded access for security. 
WARNING
Only qualifi ed service personnel should attempt to install or service this equipment. Serious injury 
may result from the improper installation or use of this equipment.