Hitachi Microscope & Magnifier S-4800 Manual De Usuario

Descargar
Página de 294
 
3 - 40
Relationship between minus tilting angle and Z position is the same as manual stage. 
 
Movable range of X and Y directions (3 axes motorized stage) 
Type1 
Specimen 
Size 
X Y 
φ
15mm 
17.5∼32.5 17.5∼32.5 
φ
25mm 
12.5∼37.5 12.5∼37.5 
φ
50mm 
0∼50 0∼50 
φ
75mm 
0∼50 0∼50 
φ
100mm 
0∼50 0∼50 
φ
125mm 
0∼50 0∼50 
φ
150mm 
0∼50 6∼50 
□100mm 0∼50 6∼50 
□125mm 0∼50 6∼50 
 
NOTICE 
When optional detector has been inserted to its operating position, further limitation is 
required.  
Refer to the operation manual of using optional detector.   
 
 
3.5.6.5  Stage Locking 
 
For high magnification work, stage locking is recommended for better mechanical stability.    The 
Z and T axes are locked or released by Lock button on the specimen stage.   
 
Release the lock to move Z and T axes. 
 
 
 
 
 
CAUTION
 
 
If the Z or T axes are moved while stage is locked, the stage mechanism may be damaged. 
X, Y and R axes are free and available when the stage is locked.