Hitachi Microscope & Magnifier S-4800 Manual De Usuario

Descargar
Página de 294
 
3 - 39
 
Important 
In case of 
φ
15mm and 
φ
25mm specimens, by applying limitation in X and Y directions written in 
page 3-40, the figure for 3 axes motorized stage (shown below) can be used. For avoiding 
damage of specimen and objective lens, be careful not to exceed limits of each axis. 
 
3.5.6.4    3 axes motorized stage (Type1 stage option) 
 
This is motorized control of type 1 stage in X, Y, R directions. Figure of relationship between 
allowable tilting angle for various Z position is shown below. For details of this option, refer to the 
manual of 3 axes motorized stage. 
0
10
20
30
40
50
60
70
0
5
10
15
20
25
30
Z (mm)
T (deg)
 φ25mm
 φ15mm
 φ50mm
 φ75mm
 φ100mm
 φ125mm
 φ150mm
CAUTION
 
 
Set Z and T within the allowable range; otherwise, the specimen may strike the objective lens 
and cause damage to the specimen and objective lens.