Hitachi Microscope & Magnifier S-4800 Manual De Usuario

Descargar
Página de 294
 
3 - 41
3.5.7    Operation of the Specimen Stage (Type 2 - 5-Axis Motorized Stage) 
 
The S-4800 type 2 SEM provides a 5-axis motorized stage 
 
3.5.7.1 Coordinate notation 
 
S-4800 utilizes two coordinate notation system, stage (mechanical) coordinate and sample 
coordinate. The sample coordinate system will be better for using the Rotation Assist function. 
The stage coordinate system is also available for users familiar with previous Hitachi SEMs using 
the stage coordinate system. 
 
(1) Relationship of stage and sample coordinates 
Following is an example of 5 inches diameter sample.   
On the stage coordinate system, the center of the sample is (X=35,Y=55mm) and the origin 
is at the upper-left corner of XY movable area.     
On the sample coordinate system, the center of the sample is (X=0,Y=0mm) and the origin 
fits to it.     
 
Stage movable area
5 inches sample
(0,0)
(110,0)
(0,110)
(110,110)
(35,55)
(0,55)
(-27.5,55)
(137.5,55)
Stage movable area
5 inches sample
(-35、-55)
(75,-55)
(-35,55)
(75,55)
(0,0)
(-35,0)
(-62.5,0)
(62.5,0)
(0,62.5)
(0,-62.5)
(35,-12.5)
(35,117.5)
X
Y
X
Y
Stage (mechanical) coordinate             Sample coordinate system 
 
(2) Selecting a coordinate system 
Open the Optional Setup dialog window by selecting Stage command in Optional Setup 
menu. 
On the COORDINATE area, select Sample or Stage. OK button will update the coordinate 
system.  
 
NOTICE 
The sample coordinate is not exactly the coordinate of the sample. It is not rotated with the 
sample when it is rotated.