3 - 60
 
                                             
Stage movable area
5 inches sample
(-35、-55)
(75,-55)
(-35,55)
(75,55)
(0,0)
(-35,0)
(-62.5,0)
(62.5,0)
(0,62.5)
(0,-62.5)
X
Y
(-34,10)
(-50,10)
A
B
 
 
The gray area is not covered with X and Y movable range but can be observed with stage 
rotation. If the point A is specified by coordinate (X=50, Y=10) input, the following message will 
be shown. 
 
                                           
 
 
Select a rotation angle and click OK. The stage will be rotated and X,Y axes are moved so as 
the specified point comes under the beam. If you click Cancel, stage will not move and input 
coordinate value is canceled. 
 
When the point B where is within movable range but is very near (less than 2mm) to X axis 
limit is specified, the following recommendation message will be shown.   
 
                                           
 
 
OK rotates the stage and move X and Y axes so as the specified point comes under the beam.   
Move without assist moves the stage without rotation. 
 
(2) Using Assist button 
When you click Assist button beside the Position display, one of following    two message will