Hitachi Microscope & Magnifier S-4800 Manual De Usuario

Descargar
Página de 294
 
3 - 62
 
The above (a) and (b) are basically dependent on in assembling error of the instrument.      These 
are measured and corrected once in the beginning (before delivery of the instrument).    But since 
the beam irradiation position may shift when adjusting the electron beam axis, this measurement 
tool is provided to enable measurement and correction at any time.     
In the other hand, the specimen height error (the above (c)) will change whenever specimens are 
exchanged. It shall be compensated at each specimen exchange to get best accuracy.   
It is not necessary when you use the eucentric rotation function with no tilting condition.   
 
(1) Specifing eucentric or non-eucentric rotation or tilt 
Check the Eucentric check box in the XYR block to enable the eucentric rotation.   
It is effective when rotate R axis by specifying the angle value. Not effective when rotate R 
axis using continuous rotation. 
 
Check the Eucentric Tilt check box in the Z/Tilt block to enable the eucentric Tilt.   
It is effective when tilting specimen by specifying the angle value. Not effective when using 
continuous tilting. 
 
(2) Calibration 
Following calibration commands are provided. 
 
(a) Measure and correct the deviation between rotation center axis and beam irradiation 
position.  
(b) Measure and correct the deviation between tilt center axis and beam irradiation position. 
(c) Measure and correct the deviation between rotation center axis and beam irradiation 
position caused by changing Z axis position.
 
(d) 
Measure the specimen height error and compensate the rotation center and the tilt center 
axes. 
 
The above (a) to (c) are basically dependent on in assembling error of the instrument.     
These are measured and corrected once in the beginning (before delivery of the instrument).   
But since the beam irradiation position may shift when adjusting the electron beam axis, this 
measurement tool is provided to enable measurement and correction at any time.     
In the other hand, the specimen height error (the above (d)) will change whenever specimens 
are exchanged. It shall be compensated at each specimen exchange to get best accuracy.   
It is not necessary when you use the eucentric rotation function with no tilting condition. 
 
To carry out the calibrations, use a flat sample and set the specimen height as close to the 
standard height as possible.   
 
 CAUTION 
The R, Z and T axes of the stage are driven automatically in this series of operations.   
So before starting operation, check the following points.