gohastings.com Hastings Entertainment Oxygen Equipment HFM-I-401 Manual De Usuario

Descargar
Página de 35
 
 
- 11 - 
2.9.  Auxiliary Input Connection 
The Hastings 400 Series flow meters provide an auxiliary analog input function.  The flow meter 
can read the analog value present between pins 5 and 6 on the terminal strip (as shown in 
Figure 2-2) and make its value available via the digital interface. The accepted electrical input 
signal is the same as that configured for the analog output signal (4 – 20 mA, 0 -20 mA, 0-5 Vdc, 
1-5 Vdc, or 0-10 Vdc).  Unlike the analog output signal, which is isolated and capable operating 
at common mode offsets of over 1000V, the analog input signal cannot be galvanically isolated 
from ground potential. 
2.10. Rotary Gas Selector 
The Hastings 400 Series flow meters can have up to eight different 
calibrations stored internally. These are referred to as gas 
records. These records are used to select different gases, but they 
can also be useful in other ways; for instance reporting the flow in 
an alternate range, flow unit or reference temperature. The 
records are referred to by their number label from #0 – #7.  
 
The first six records will, by default, be setup for most common 
six gases as shown in Figure 2-11. If a gas other than one of these 
six is specified on the customer order it will be placed in record 
#6. If a second different gas is selected, it will be placed in record 
#7. If multiple different gases or ranges are specified they will 
replace some of the standard six gases.  
 
The purchased calibration certificate is provided for the gas (or gases) specified by the 
customer when ordering. This gas will be indicated with an “X” on the Gas Label (diagram 
below) that is located on the top of the 400 Series Mass flow meter’s electronics enclosure.  
The remaining gas records will have a different full scale value and an unverified calibration. 
The full scale range can be calculated by using the Gas conversion factor or GCF.  A 
comprehensive list is found in Appendix 2 in this manual. 
 
 
Example 1 
To convert the calibration of a full scale range of 100 slm of Nitrogen to 
the other full 
scale ranges: 
 
 
1. Calculate full scale value of Helium 
Calibrated gas = Nitrogen (GCF
1
 = 1.000) 
Full scale range (FS
1
) = 100 slm 
Secondary gas (FS
2
) = Helium (GCF
2
 = 1.40) 
1.40/1 = 1.40, 1.40 x 100 = 140 slm of Helium 
 
2. Calculate full scale value of Hydrogen 
Record# Gas 
0 Nitrogen 
1 Air 
2 Helium 
3 Hydrogen
4 Argon 
5 Oxygen 
6 Custom 
7 Custom 
Figure 2-11 Gas record table 
Record# Gas 
Full S
Rang
0 Nitrogen 
100 slm 
1 Air 
100.15 slm 
2 Helium 140 slm 
3 Hydrogen 100.38 slm 
4 Argon 140.37 slm 
5 Oxygen 
97.95 slm  
6 Custom 
Not included i
specified 
7 Custom 
Not included i
specified 
1
2
1
2
GCF
GCF
FS
FS
=
0
1
2
3
N
Air
He
H
4
5
6
Ar
2
O
2
7
X= cal report 
generated
(others use GCF)
S/N
2
X