Emerson Process Management 2400S Manuel D’Utilisation

Page de 208
26
Micro Motion
®
 Model 2400S Transmitters for PROFIBUS-DP
Required Transmitter Configuration
Figure 6-1
Sample calibration tags 
Flow calibration values
Two factors are used to define flow calibration:
The flow calibration factor, which is a 6-character string (five numbers and a decimal point)
The temperature coefficient for flow, which is a 4-character string (three numbers and a 
decimal point)
These values are concatenated on the sensor tag, but different labels are used for different sensors. As 
shown in Figure 6-1:
For T-Series sensors, the value is called the FCF value. 
For other sensors, the value is called the Flow Cal value.
Table 6-1
Sensor calibration parameters
Parameter
Sensor type
T-Series
Other
K1
K2
FD
D1
D2
Temp coeff (DT)
(1)
(1) On some sensor tags, shown as TC.
Flowcal
(2)
(2) See the section entitled “Flow calibration values.
FCF
FTG
FFQ
DTG
DFQ1
DFQ2
Other sensors
19.0005.13
0.0010
0.9980
12502.000
14282.000
4.44000
310
12500142864.44
T-Series