Emerson 3812 Manuel D’Utilisation

Page de 138
Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual
Section 2: Mechanical installation
3-9000-765 Rev H
May 2015
Use of appropriate safety engineered swivel hoist rings in meter end flanges
 35
8.
Always use separate slings  for each hoist ring. NEVER reeve one sling through both 
hoist rings. The slings must be of equal length. Each sling must have a load rating that 
equals or exceeds the hoist ring load rating. The angle between the two slings going to 
the hoist rings must never exceed 90 degrees or the load rating of the hoist rings will be 
exceeded.
Figure 2-9  90 Degree angle between slings