Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual
Section 2: Mechanical installation
3-9000-765 Rev H
May 2015
Use of appropriate safety engineered swivel hoist rings in meter end flanges
 37
10.
Remote mount option: Always use separate slings  for each hoist ring. NEVER reeve one 
sling through both hoist rings. The slings must be of equal length. Each sling must have 
a load rating that equals or exceeds the hoist ring load rating. The angle between the 
two slings going to the hoist rings must never exceed 90 degrees or the load rating of 
the hoist rings will be exceeded. 
DO NOT DRAG THE TRANSMITTER ELECETRONICS ENCLOSURE DURING LIFTING 
OPERATIONS.
Support the transmitter electronics while lifting the meter body.
Failure to do so may r cause damage to the equipment.
REMOVE CONDUIT TIE WRAPS FROM THE JUNCTION BOX 
Conduit tie wraps must be removed prior to powering the meter.
Tie wraps placed on the junction box conduit for protection during shipping must be 
removed before the meter is powered.