Invacare 1150704 Manuel D’Utilisation

Page de 68
SECTION 3—OPERATION
13
Activating a Mechanical Emergency Release
Primary Emergency Release
NOTE: For this procedure, refer to FIGURE 3.2.
NOTE: This procedure will bring the boom down or lift it up if the hand 
control is not functioning properly.
To activate the primary emergency release, insert a pen into the 
hole labeled Emergency Up or Emergency Down on the control 
box.
Secondary Emergency Release
NOTE: For this procedure, refer to FIGURE 3.3.
NOTE: All patient lift actuators are equipped with a mechanical 
emergency release. The mechanical release will enable the actuator to 
retract without power.
NOTE: Use the primary emergency release first before using the 
secondary emergency release procedure. This procedure should only be 
used if the primary emergency release procedure is not functioning or 
is unreachable.
NOTE: The lift MUST be under a load for the mechanical release to 
function.
To activate the secondary emergency release, pull up on the RED 
emergency grip and pull down on the boom at the same time.
Charging the Battery
NOTE: For this procedure, refer to FIGURE 3.4.
NOTE: Invacare recommends the battery be recharged daily to prolong 
battery life. 
NOTE: An audible alarm will sound when battery is low.
1.
Lift up on the handle on the back of the battery.
2.
Lift the battery up and away from the control box.
CAUTION
Make sure there is an audible click when mount-
ing battery on the battery charger to ensure 
proper mounting. Otherwise, injury or damage 
may occur.
3.
Place the battery on the battery charger. Push the top of the 
battery against the mounting bracket until there is an 
audible click.
NOTE: The charge LED will illuminate. When charged, the LED will 
stop illuminating.
NOTE: It will take approximately four hours to charge a battery that 
requires a full charge.
4.
Lift up on the handle on the back of the battery.
5.
Lift the battery up and away from the battery charger.
CAUTION
Make sure there is an audible click when mount-
ing battery onto the control box to ensure proper 
mounting. Otherwise, injury or damage may 
occur.
6.
Place the battery on the control box. Push the top of the 
battery against the mounting bracket until there is an 
audible click.
FIGURE 3.2   Primary Emergency Release
FIGURE 3.3   Secondary Emergency Release
FIGURE 3.4   Charging the Battery
Control Box
Emergency Down 
Hole
Emergency Up 
Hole
RED
Emergency Grip
Battery Charger 
(STEP 3)
Control Box 
(STEP 6)
An audible “click” 
will be heard when 
properly installed
(STEPS 3 and 6) 
Handle
(STEPS 1, 2, 4 and 5)
Battery