Description
Fault
Actions when fault is detected:
The alarm is posted to the Alert List.
Outputs go to the configured fault action (after Fault Timeout has expired, if
applicable).
Digital communications go to the configured fault action (after Fault Timeout
has expired, if applicable).
The status LED (if available) changes to red or yellow (depending on alarm se-
verity).
Actions when alarm clears:
Outputs return to normal behavior.
Digital communications return to normal behavior.
The status LED (if available) returns to green and may or may not flash.
Informa-
tional
Actions when fault is detected:
The alarm is posted to the Alert List.
The status LED (if available) changes to red or yellow (depending on alarm se-
verity).
Actions when alarm clears:
The status LED (if available) returns to green and may or may not flash.
Ignore
No action
Status alarms and options for Status Alarm Severity
Status alarms and Status Alarm Severity
Table 5-2:   
Alarm code
Status message
Default severity Notes
Configurable?
A001
EEPROM Error (Core Pro-
cessor)
Fault
No
A002
RAM Error (Core Processor) Fault
No
A003
No Sensor Response
Fault
Yes
A004
Temperature Overrange
Fault
No
A005
Mass Flow Rate Overrange Fault
Yes
A006
Characterization Required
Fault
Yes
A008
Density Overrange
Fault
Yes
A009
Transmitter Initializing/
Warming Up
Fault
Yes
A010
Calibration Failure
Fault
No
A011
Zero Calibration Failed:
Low
Fault
Yes
A012
Zero Calibration Failed:
High
Fault
Yes
A013
Zero Calibration Failed:
Unstable
Fault
Yes
Configure device options and preferences
Configuration and Use Manual 
73