SITRANS LU01 – INSTRUCTION MANUAL
Page 49
m
mmmm
Pa
ra
m
e
te
rs
Preset is determined by mA Function (P200). If P200 = 1 or 3, then the preset is 0.0, or if 
P200 = 2 or 4, then the preset is 3.8. For P200=1 or 3 (preset is 0.0), this parameter has no 
effect because the minimum limit cannot be negative, and the minimum current is always 
0.0 mA.
P213 mA Output Max Limit 
Sets the maximum mA output value (in mA) to be produced.
mA Output Trim (P214 to P215)
This does not affect the P203 value shown, and is used when recalibration of an external 
device is impractical.
P214 4 mA Output Trim 
Calibrates the 4 mA output.
Adjust this value so the device indicates 4.000 mA when P214 is accessed. 
P215 20 mA Output Trim 
Calibrates the 20 mA output.
Adjust this value so the device indicates 20.00 mA when P215 is accessed.
Primary Index
mA output
Values
Range: 0.000 to 22.00
Preset: 20.2 mA
Related
P200 mA Output Range / P212 mA Output Min LImit
Primary Index
mA output
Values
Range: -1.00 to 1.000
Preset: 0.000
Related
P215 20 mA Output Trim
Primary Index
mA output
Values
Range: -1.00 to 1.000
Preset: 0.000
Related
P214 4 mA Output Trim