P302 Temperature, Sensor Maximum
View the highest temperature encountered, as measured by the TS-3 Temperature 
Sensor (if applicable).
Press CLEAR 
 keys to reset the log after a short circuit on the transducer wiring.
Profile Records (P330 to P337) 
These features can record up to ten Echo profiles, initiated manually (P330), or 
automatically (P331 et al) for viewing at a later time using Dolphin Plus or an oscilloscope. 
See 
Scope Displays
 (P810) for echo profile viewing hardware / software requirements. If 
ten Profiles are already saved, addresses 1 through 10 are filled, the oldest automatically 
initiated record is overwritten. Manually initiated records are not automatically 
overwritten. All records are automatically deleted in the event of a power interruption.
When a record is displayed, results are based on current programming (which may have 
been altered since the record was saved). This permits the effect on the echo profile to be 
observed when changing an echo parameter.
P330 Profile Record
Records profiles for later viewing.
In addition to being a profile records library, this provides two functions:
manually records and saves echo profiles
displays an echo profile, recorded manually or automatically (oscilloscope)
To select a record address
1.
Enter PROGRAM mode and press DISPLAY 
 twice to highlight the index field.
The field shows two underscores _ _.
2.
Type the index number. The profile record information is shown. 
3.
Use ARROWS 
 
 to scroll through the records.
Values
Range: - 50 to 150°C (view only)
Preset: - 50°C
WARNING: 
These parameters are for authorized service personnel or 
technicians familiar with Siemens Milltronics echo processing techniques.
Primary Index
Echo profile
Values
Code
Description
----
no record
A1
automatically recorded profile from Transducer One
U1
manually recorded profile from Transducer One