Emerson Network Router Manuel D’Utilisation

Page de 110
98
ProLink
®
 II Software for Micro Motion
®
 Transmitters
Configuring the Discrete Batch Application
B.5
Performing Batch AOC calibration
Batch AOC (Automatic Overshoot Compensation) keeps the actual quantity delivered as close as 
possible to the batch target, by minimizing the amount of overshoot.
If batch AOC is enabled (see Section B.3.2), batch AOC calibration is required to provide data for the 
compensation process. Thereafter, batch AOC calibration is recommended:
Whenever consistent overshooting or undershooting of the target occurs
If equipment (valve or pump) is changed
Note: If batch AOC is not enabled, you cannot perform batch AOC calibration.
To perform batch AOC calibration using ProLink II:
1. In the 
File
 menu, click on 
Calibration
.
2. Click on 
AOC calibration
.
3. Press the 
Start Calibration
 button. The 
AOC Calibration Active
 light turns red, and will 
remain red while batch AOC calibration is in progress.
4. Run 2–10 batches.
5. When overshoot has been adequately minimized, press the 
Save Calibration
 button.