5.5.3
Temperature calibration
Temperature calibration is a two-point procedure: temperature offset calibration and temperature 
slope calibration. The entire procedure must be completed without interruption. To perform a 
temperature calibration, see Figure 5-7.
Figure 5-7
Temperature calibration
Fill sensor with 
low-temperature fluid
Fill sensor with 
high-temperature fluid
Calibration
Temp slope cal
Do Cal
Enter temperature of 
high-temperature fluid
Calibration in Progress 
light turns red
ProLink
Calibration in Progress 
light turns green
Wait until sensor achieves 
thermal equilibrium
Calibration
Temp offset cal
Do Cal
Enter temperature of 
low-temperature fluid
Calibration in Progress 
light turns red
ProLink
Calibration in Progress 
light turns green
Wait until sensor achieves 
thermal equilibrium
Close
Close