Terminating resistor
When connecting several appli-
ances via the RS485 interface, the 
terminating resistor must remain 
on the last appliance.
Remove the terminating resis-
tors from the appliances in 
between.
External alarm
We recommend connecting the appliance 
to an external alarm device.
There are various connection options at 
the back of the appliance.
The appliance may only be connected 
to an external alarm device by trained 
personnel.
Floating alarm 
output
Connection to 
RS485 interface
Note
The connectors are secured with 
screws. To remove the connectors, 
undo the left and right screws.
Optional temperature 
sensor connection
RS485 interface
Rx- / Tx- 
Send/Receive data cable (negative 
pole)
Rx+ / Tx+ 
Send/Receive data cable (positive 
pole)
GND
Earth cable
N.O 
Alarm output
Connection for a visual warning light or 
an acoustic alarm signal.
N.C
Operating light
Connection for a control lamp to indicate 
that the appliance is in normal mode.
COM
External power supply unit
42 V/8 A DC maximum.
Minimum current: 150 mA.
Floating alarm output
These three contacts can be used to connect the appliance to an 
optical or acoustic alarm device.
The connection is designed for a maximum of 42 V/8 A DC from 
a safety extra-low voltage (SELV) source (minimum  current: 
150 mA).
Important
When supplying mains voltage to the floating alarm contact, 
the technical safety requirements of standard EN 60335 will 
not be satisfied.