Tektronix TVS600 & TVS600A Series Manuel D’Utilisation

Page de 410
SYSTem Subsystem
2–188
TVS600 & TVS600A Command Reference
SYSTem:BDATe?
SYSTem:BDATe
TVS600A Models Only
Sets and queries the battery installation date for the nonvolatile memory in the
instruments.
The date loaded is not evaluated for accuracy by the instrument. However the
year should be the complete four digit year ie. 1997 and the month should be 1
through 12 for January through December and the day should be 1 through 31
for the day of the month.
SYSTem:BDATe?
SYSTem:BDATe <year>,<month>,<day>
<year> <month> <day>
Query response
<NRf>, <NRf>, <NRf>
<NR1>,<NR1>,<NR1>
None
Not applicable
Execution Error –203, “Command protected”
Attempted to set battery date when SYSTem:PROTect is ON.
None
Command:
SYST:BDAT 1997,7,4
Query:
SYST:BDAT?
Response:
1997,7,4
SYSTem:PROTect
Syntax
Parameters
Suffix Units and
Multipliers
Reset Value
Errors and Events
Dependencies
Examples
Related Commands