TRIGger:PULSe Subsystem
TVS600 & TVS600A Command Reference
2–271
TRIGger:PULSe:GLITch:POLarity
TRIGger:PULSe:GLITch:POLarity?
Sets or queries the polarity of the event pulse for pulse glitch triggering. Before
this command will be effective, you must set TRIGger:TYPE to PULSe and
TRIGger:PULSe:CLASs to GLITch. Refer to the discussion of TRIG-
ger:PULSe:CLASs on page 2–270.
TRIGger[:A]:PULSe:GLITch:POLarity <polarity>
TRIGger[:A]:PULSe:GLITch:POLarity?
<polarity>
Query response
EITHer
NEGative
POSitive
EITH
NEG
POS
POS
Execution Error –224, “Illegal parameter value”
Attempted to set polarity to an illegal value.
None
Command:
TRIG:PULS:GLIT:POL NEG
Query:
TRIG:PULS:GLIT:POL?
Response:
NEG