AeroScout FM Manuel D’Utilisation
Hardware Features
4
M310 Fall Monitor – Setup & User Guide
Speaker Opening
The speaker opening allows the monitor alarm to be
audible.
audible.
Reset Button
The reset button is used to display the monitor’s status
without disrupting the monitoring session, to mute
audible alerts, or to put the monitor into Hold Mode.
Hold Mode is used to suspend patient monitoring to
allow staff to move the patient temporarily (adjust
bedding, trip to the toilet, etc.)
Button press duration variations are as follows:
• A short press is used to display the monitor’s status or cancel any
without disrupting the monitoring session, to mute
audible alerts, or to put the monitor into Hold Mode.
Hold Mode is used to suspend patient monitoring to
allow staff to move the patient temporarily (adjust
bedding, trip to the toilet, etc.)
Button press duration variations are as follows:
• A short press is used to display the monitor’s status or cancel any
audible alarms.
• A long press, more than 3 seconds, is used to place the monitor in
Hold Mode for 2 minutes (120 seconds).
Status Indicators
Pad
This LED indicates when a pad is
disconnected, or when two pads are
being monitored simultaneously. PAD
icon flashes, and honk alert sounds
every 5 seconds while 2 pads are being
monitored.
disconnected, or when two pads are
being monitored simultaneously. PAD
icon flashes, and honk alert sounds
every 5 seconds while 2 pads are being
monitored.
Nurse Call
Indicates when the monitor is set to remote alert using the Nurse Call
system but the cable is unplugged.
system but the cable is unplugged.
Battery
This LED flashes YELLOW on low battery, or steady GREEN with
YELLOW flashes when connected to a power supply and batteries are
low or not installed.
YELLOW flashes when connected to a power supply and batteries are
low or not installed.