button or the Exit button. For more information about these options, 
refer to the System Guide for the Guidant pulse generator being 
interrogated.
6.
To proceed with an interrogation session or read data from a patient 
data disk (which is a feature that is available in some applications), 
refer to the System Guide for the Guidant pulse generator being 
interrogated.
Changing Parameter Values
The screens for many of the features contain parameter information, 
including parameter names, values for the present parameter settings, 
value boxes to accommodate value changes, and buttons to cancel 
changes. Many screens display two columns next to the parameter 
names: the Present column and the Change column. The Present column 
displays the currently programmed value for each parameter, and the 
Change column displays value boxes that allow the clinician to make 
changes to the parameters. If a particular parameter is not applicable in 
the current mode, dashes will be displayed in the Present column. If a 
particular parameter is not applicable to the mode in the Change column, 
that parameter’s value box is not displayed.
Palette Window: To change the parameter values, first select the 
appropriate parameter’s value box in the Change column. A palette 
window will appear (as shown in Figure 8). Select a value from the palette 
window (by touching the desired value and lifting the stylus from the 
screen). Touching the screen outside a window will close it without 
making a selection. When changes have been made to parameter values, 
the data appear in the Change box until programmed into the pulse 
generator.
In some screens (e.g., Patient Data screens), data are not displayed in 
Present and Change columns. These screens display value boxes that 
Figure 8. Types of windows enabling value changes.
Palette Window
Keyboard Window