Encompass 4 Reader System Guide
7-18
O1
RF on
T1
Uniqueness timeout of two minutes
T2
Uniqueness timeout of 15 seconds
T3
Uniqueness timeout of 30 seconds
Fxxx
RF output frequency, xxx = 000 to 118 hexadecimal offset in 250 
kHz from 860 MHz. If an invalid frequency value is stored 
(corrupted NVRAM), then xxx = “XXX” to indicate an error in 
the frequency setting.
Rxx
Tag decoder range (distance) for ATA tags, xx = 00 to 1F 
hexadecimal range value 
Gxx
Tag decoder range (distance) for SeGo protocol tags, xx = 00 to 
1F hexadecimal range value
Axx
RF power attenuation, where 00 is maximum output power and 
0A is minimum output power (10dB less than maximum power).
Ixx
IAG RF power attenuation, where 00 is maximum output power 
and 0F is minimum output power (15 dB less than maximum 
power)
For example, if factory default settings are assigned, the reader message is
RFST C1 O0 T1 Fxx R1F G1F A00 I04
which means that RF is enabled by presence sensor on input0, RF signal off, unique-
ness timeout of two minutes, RF output frequency has not been set, maximum RF out-
put range for ATA tags, maximum RF output range for eGo and eGo Plus tags, full RF 
power, and IAG power set at 4 dB attenuation.
529 Display Presence Input Status
Command #529 displays the parameters associated with presence detection and RF 
control. The reader’s message indicates if presence without tag reports are enabled/
disabled (#690N Select Presence without Tag Report Option), if input inversion is 
enabled/disabled (#694N Select Input Inversion Option), and the minimum presence 
true period (always true). The reader’s message also reports the selected RF timeout 
(#693N Select RF Timeout Period) and the selected means of RF-off control (#692N 
Select RF Control Algorithm). If presence without tag reports is enabled (#6901 
Enable Presence without Tag Reports), the reader transmits a report if a presence is 
detected without the subsequent acquisition of a valid tag.
Note:  RF timeout values vary depending on the operative tag read mode and the type 
of tag in the read field. All times are approximate.
Reader response:
PRST P<0 to 1> D0 A<0 to 2> T<0 to F> I<0 to 1>