Setup for PC software
57
Enter your PIN and click OK. This opens the Setup Parameters dialog box.
Type the desired institution name in the space provided and press OK to accept or Cancel to 
abort.
To log active pods
When the system is powered up, the connected Pods are not communicating with the software. 
To establish communication, you must activate the Log Active Pods command on the Setup menu.
Connect and power-up the system.
Select Setup > Log Active Pods.
Wait approximately 15 seconds for the connected Pods to appear on the main screen. 
NOTE: If the Pods do not appear, check to be sure the battery switches are on.
To check the battery
NOTE: Only Model 1136 rf-IVD
2
 wireless Pods have batteries.
You can check the remaining battery life and the state of each battery from the display on the Dose 
view (Figure 4-19). 
Figure 4-19. Battery status (Bat)
Battery status is shown as follows:
Three choices are selectable in parameter setup screen: (1) Percent remaining Bar, (2) 
Percent remaining Text, and (3) Hours:minutes remaining
If hours and minutes are shown, the numbers displayed are predicted hours and minutes 
of measurement time left (not including a 30-minute safety factor). If the system is idle 
(not measuring), the batteries will last about twice as long as shown on the display. See 
“About predicted battery life” on page 87.
Status: c = charging (battery is discharged and is charging at maximum rate), t = trickle 
charging (a full charge is being maintained with a trickle of power), d = discharging, o = 
battery switch is off. 
A battery can show “off” when fully charged.