Sun Nuclear Corporation 1136 Manuel D’Utilisation

Page de 98
Setup for PC software
55
Connecting multiple instruments to USB
If you use multiple instruments with one computer (for example, an IVD and a DQAC) each instal-
lation is assigned a different serial port number. You must know which serial port is assigned to 
each instrument and select only that port when running that instrument. Or you can use the 
“Detect” or “Find Port” commands to detect the instrument.
Removing the USB drivers
To remove the USB drivers, open Control Panel (Start > Settings > Control Panel) and double click 
on Add/Remove Programs. In the Add/Remove Programs dialog box, select “FTDI USB Serial Con-
verter Drivers” and click Change/Remove. Follow the on-screen directions.
Parameter setup
The Parameter Setup item on the Setup menu lets you 
enter or change the PIN (Personal Identification Number), 
select the printout characteristics
select the type of display for battery status,
select the type of dose units to be displayed, 
set the RV Portal Software to run at start, 
suppress background subtraction for uncalibrated inputs, and 
enter an Institution Name that will be printed at the top of the printed reports.
To set or change a PIN 
The administrator PIN protects patient safety by preventing unsafe connections. In addition, the 
PIN avoids accidental changes to setup and calibration. To use no PIN, enter 0000. 
Select Parameter Setup from the Setup menu. This opens the Setup Parameters dialog box 
(Figure 4-18).
Figure 4-18. Setup parameters dialog box
Click the Setup Physics Pin button to open the Password dialog box.
In the Password dialog box, enter the PIN once, then enter it again to confirm.
Click OK.
NOTE: If the PIN is lost or forgotten, the physicist can key in “0911” to display the stored PIN. After 
0911 is entered, fill in the stored PIN and press ENTER to continue. Then the existing PIN can be 
used, or the PIN can be changed.
About the medical power supply
The medical power supply, P/N 741008, is required for use with the Models 1136 and 1137. The 
medical power supply can be identified by the label “Medical Power Supply” on the top surface.