Beha Amprobe GT-800 PRO KITVDE-tester 4472081 Manuel D’Utilisation

Codes de produits
4472081
Page de 76
GT-600/GT-800 Instruction Manual 
16 
 
 
Limit value setting 
 
Limit value is offered in all functions except in P/I
L
 and CLAMP. For limit value 
setting press the “LIM” key (soft menu key F4) then use the soft menu keys “+”, 
“–“, “STD” (Standard) or “CALC” (Calculation). The soft menu key F8 “EXIT” can  
be used to exit limit value setting display.  
  In case that test result comply with set limit value, the mark “ ” is 
displayed and two short sound signals are present after finishing the 
measurement. In case of non-compliance, the mark “X” is displayed and 
one longer sound signal is present after finishing the measurement. The 
limit value is saved as a parameter of the measurement result and is 
transferred to PC in case of data transfer. 
 
Start of the measurement 
 
  For single measurement press the START key and release it. In this case 
measurement time is defined in MENU/SETUP/MEASUREMENT TIMES menu, 
see the chapter “SETUP menu” on page 49. Watch elapsed time on displayed 
bar graph during the measurement.  
  For continuous measurements press and keep pressing the START key for at 
least 2 seconds until “CONT” (continuous) information is displayed. 
Measurement time is limited to 5 minutes in this case. Continuous 
measurement is available in all functions except in CORD, PRCD and AUTO 
test. 
 
 
External Voltage Display 
 
  If there is a dangerous external voltage present on test terminals prior to a 
test or during it, the warning message “EXTERNAL VOLTAGE!“ will appear on 
the display and the start of the measurement is blocked. Remove the external 
voltage! 
  If an external voltage is applied to test terminals during the test in R
ISO
 
function, wrong measurement values may be displayed. 
      
                                                                                      
 
 
 
Testing Appliances 
 
 WARNINGS 
  Before commencing any testing you are strongly advised to make reference to 
the local regulations and standards for safety at works regulations and any 
relevant publications from the Health and Safety Executive. 
  The appliance must be switched on for all tests. 
  When conducting tests do not touch the appliance as some tests involve high 
voltages and high currents. 
 
 
Please remove external voltage from any test terminal 
immediately if “ EXTERNAL VOLTAGE!” warning is displayed.