Protective Bonding Test 5 A / 0.2 A (R
PE
 
The protective bonding test measures the resistance between PE terminal of test 
socket and PROBE 1 terminal. This test applies to class I appliances only. 
 
Remarks: 
  To obtain correct protective bonding test results you must have 
compensated (zeroed) the test lead, see the section “Test lead 
compensation” on page 15. 
  Connect the appliance and the earth bond test lead according to the figure 6 
or 7. 
  You should use the lower test current 0.2 A for certain appliances. Please 
refer to local appliance test standards and guidance material. 
  During the measurement flex the flexible cord along its length to help find 
any broken conductors or poor quality joints. 
 
 
 
Figure 6: Protective bonding test connections at appliance with mains plug 
 
 
 
 
Figure 7: Protective bonding test connections at permanently connected  
  
appliance