GT-600/GT-800 Instruction Manual
40
Protective Extra Low Voltage Test (PELV)
The test measures the protective extra low voltage between any accessible
conductive part of tested appliance and earth (PE).
conductive part of tested appliance and earth (PE).
Figure 32: PELV test connection
Display
explanation:
Figure 33: Display in PELV test
1 ...... “SELV/PELV” soft menu key, for selection of measurement function
(SELV or PELV).
2 ...... “LIM” soft menu key, for pass/fail limit value setting (standard values
are 25 or 50 V).
3 ...... Test result status ( ... result OK, X ... result not OK).
4 ...... Pre-set pass/fail limit value.
5 ...... Function selected by the rotary switch and the “SELV/PELV” key.
6 ...... Test result and unit.
4 ...... Pre-set pass/fail limit value.
5 ...... Function selected by the rotary switch and the “SELV/PELV” key.
6 ...... Test result and unit.
Specific information that can be shown on the display
Information displayed
Description
<10.0V
PELV value lower than 10.0 V.
>150.0V
PELV value higher than 150.0 V. Remove the
voltage.
voltage.
In order to save displayed measurement result, press the “SAVE” function
key (F5) twice, for further instructions see the “Memorizing example”
section.
section.